Oxygen made on site, at 93 ± 3 %.

UNISSI installs PSA oxygen generator plants, restores plants that have stopped, and runs and maintains them under CAMC with operators on site. We work on plants from 160 to 1,500 LPM, including PM CARES and DRDO installations.

ABO₂

How a PSA plant works

Air in, 93 % oxygen out.

A PSA plant makes oxygen by pushing dry, clean air through zeolite that holds back nitrogen. Two towers take turns: one makes oxygen while the other vents the nitrogen it has collected. Tap any part of the plant to see what it does and what we check on a CAMC visit.

Situation
  1. A makes O₂ · B vents
  2. Equalise
  3. B makes O₂ · A vents
  4. Equalise

Swipe sideways to see the whole plant.

PSA oxygen plant schematic Air is compressed, stored, filtered and dried, then fed to one of two zeolite towers. The tower under pressure holds back nitrogen and passes oxygen to the oxygen receiver, while the other tower vents its nitrogen through a silencer. An analyser checks purity before the oxygen goes to the hospital pipeline, and a booster can fill cylinders at 150 bar. A reserve manifold takes over if purity falls below 90 percent. 7.0 bar AMBIENT AIR AIR COMPRESSOR 7.0 bar AIR RECEIVER PRE-FILTER PDP +3 °C REFRIGERATED DRYER CARBON FILTER ZEOLITE TOWERS SILENCER N₂ EXHAUST 4.6 bar O₂ RECEIVER 94.5 % ANALYSER RESERVE MANIFOLD TO HOSPITAL MGPS · 4 BAR — LPM · 93 ± 3 % O₂ BOOSTER off CYLINDER BANK

Running normally. Tower A is making oxygen while tower B vents nitrogen.

Oxygen purity
94.5 %
Plant output
700 LPM
Dryer dew point
+3 °C
Hospital supplied from
PSA plant

Selected part

Zeolite towers

Typical figures

What we check on a CAMC visit

PSA plant sizing

How big a plant does your hospital need?

Enter your bed mix to get a first estimate of oxygen demand and the plant size to match. Our site survey confirms the final number.

Beds with oxygen
Planning assumptions
Running cost figures
Recommended plant—
Connected load
—
Typical demand
—
With headroom
—
D-type cylinders it replaces
—
Running cost, per month
Same oxygen from cylinders
—
Plant electricity
—
Saving
—
Payback on the plant
Add the plant cost

A planning estimate only. Flow per bed follows common planning figures; your clinical team and our site survey set the final capacity. Cylinder figure assumes about 7,000 litres of gas per D-type cylinder. Running costs use typical demand and leave out CAMC, operators and spares; change the figures under “Running cost figures” to match your hospital.

Ask for a site survey

Plant down?

First checks before our engineer arrives.

Pick what you are seeing. These are checks a plant operator or the hospital's biomedical engineer can make safely. Send us what you find and we will tell you the next step.

Patients first. If purity or pressure is low, make sure supply has switched to the cylinder manifold or LMO tank before anything else. Do not open valves, filter housings or vessels that are under pressure.

  1. Note the analyser reading, the flow on the flow meter and the O₂ receiver pressure.
  2. Check whether the hospital is drawing more than the plant's rated LPM. If it is, purity falls; bring flow down to the rating and watch whether purity recovers over 15 to 20 minutes.
  3. Check the air dryer is running and its auto-drain is discharging water. Wet air is the most common cause and damages the zeolite.
  4. Watch the two tower pressure gauges. They should swing each time the plant switches beds. A gauge that stays still points to a stuck switching valve.
Call +91 93100 31900

Inside a PSA plant

What we install, service and repair.

Every stage of the plant, with the air treatment that keeps the zeolite alive.

FEED

Air compression

  • Rotary screw compressor, 7.5 to 10 bar
  • Heavy-duty intake filters
  • Variable speed drive where specified
TREATMENT

Drying and filtration

  • Water separator
  • Refrigerated air dryer
  • 1 micron, 0.01 micron and activated carbon filters
  • Air buffer tank
CORE

Oxygen generator

  • Twin adsorber towers
  • Zeolite molecular sieve
  • Pneumatic switching valves and silencer
  • PLC control panel
  • Oxygen analyser
OUTPUT

Storage and delivery

  • Oxygen surge tank
  • Bacterial filter
  • Flow meter
  • Tie-in to the hospital MGPS
FILLING

Cylinder filling (optional)

  • High-pressure oxygen booster to 150 bar
  • Cylinder filling ramp
SAFETY

Power and safety

  • Servo stabilizers up to 150 kVA
  • Automatic changeover to DG power
  • Automatic changeover to cylinder manifold
  • Safety relief and auto-drain valves
  • Plant-room ventilation and ducting

Recent oxygen plant work

Plants we have restored and run.

  • SKIMS MCH, Bemina, SrinagarZeolite replacement, 750 LPM plant (2026)
  • GMC & AH, RajouriRestoration of 1,000 LPM PM CARES DRDO plant
  • District Hospital, RambanRepair of skid II, 1,500 LPM plant
  • District Hospital, KishtwarSkid repair, 1,500 LPM plant; plant operation
  • GMC KathuaOperation of 960 LPM plant
  • NHM AssamRepair and CAMC of plants at 29 district and sub-divisional hospitals
  • Haflong Civil Hospital, Assam150 kVA servo stabilizer for the oxygen plant (2026)
  • Military Hospital, NainitalOxygen plant CMC

Questions hospitals ask us

Oxygen plant questions

Do you maintain oxygen plants that another company installed?

Yes. Most of the PSA oxygen plants under our CAMC were installed by other firms, including PM CARES and DRDO plants and L&T and Ozcan makes, with capacities from 160 to 1,500 LPM.

What is included in a PSA oxygen plant CAMC?

Preventive maintenance visits, breakdown repairs and spares. Where the contract asks for it, we also station trained operators at the plant.

What oxygen purity does a PSA plant produce?

Medical PSA plants deliver 93 ± 3 % oxygen. An oxygen analyser monitors purity continuously, and if it falls below 90 % the supply switches automatically to the cylinder manifold.

When does the zeolite in a PSA plant need replacing?

When the plant can no longer hold purity at its rated flow even after the compressor, dryer and filters are serviced. Oil or moisture reaching the sieve shortens its life, so air treatment is checked on every visit.

What size of PSA oxygen plant does a hospital need?

It depends on the number of ICU, HDU and general beds with oxygen and on how many operation theatres the plant feeds. Use the calculator on this page for a first estimate, then ask us for a site survey.

Contact

Plant down, or planning a new one?

Send us the plant capacity, make and location. We will come back with a survey date or a quotation.

Registered office · Gurugram

WW10, Windchants, Experion, Sector 112,
New Palam Vihar, Gurugram, Haryana 122017

Delhi

Plot No 20, Village Raghopur,
Najafgarh, New Delhi 110071

Jaipur

Khasra No 53, Rajpura Patalwas,
Bassi, Jaipur, Rajasthan 303305